Detection Systems at the HTSKC (CCSR 0133 and ChEM-H W008)

Plate Readers at the HTSKC Schedule Data Sheet
BMG ClarioStarPlus w/Stackers (CCSR-0133) ClarioStarPlus (View Only) ClarioStarPlus
BMG ClarioStarPlus (ChEM-H W008) ClarioStarPlus (View Only) ClarioStarPlus
Tecan Infinite M1000-multimode plate reader (CCSR-0133) Tecan Infinite M1000 (View Only) Tecan Infinite M1000

Infinite M1000 Instructions

 

iControl V2.0 Manual
Tecan Infinite M1000 PRO-multimode plate reader w/Stackers (CCSR-0133) Tecan Infinite M1000 PRO (View Only) Tecan Infinite M1000 PRO

Infinite M1000 PRO Instructions

 

Magellan 7.1 Manual
FLEXstation II-384-fluorescent plate reader (CCSR-0133) Flexstation II 384 (View Only) Flexstation
 

 

SoftMax Pro 5 Manual
AnalystGT-multimode plate reader (CCSR-0128) AnalystGT (View Only) Analyst GT
QuantStudio 12K Flex RT PCR (CCSR-0128) QuantStudio(Genomics Core) Brochure

CLARIOstar Plus (BMG) ChEM-H W008 & CCSR-0133 (w/Stackers)

CLARIOstar Plus

 The ClarioThe BMG CLARIOstarPlus isthe HTSKC's most flexible multi-mode plate reader. It comes equipped with Linear Variable Filter Monochromators, filters, and spectrometer. It has the best sensitivity using Enhanced Dynamic Range technology and can read 6 to 1536-well microplate and can measure;

  • Absorbance (220 nm-1,000 nm)
  • Fluorescence Intensity (Ex 240nm-840 nm, Em 280 nm-900 nm)
  • Fluorescence Polarization (FP, Filter-based, any fluorophore excited by any filter listed below)
  • Time-resolved fluorescence (TRF, Filter-based)
  • Fluorescence resonance energy transfer (FRET, Filter-based)
  • Homogeneous Time-Resolved Fluorescence (HTRF, Filter-based)
  • Glow/Flash Chemiluminescence (320 - 740 nm) including BRET.
  • AlphaScreen/AlphaLISA-only in CCSR
  • Software data acquisition and analysis with Multi-user Reader Control and MARS Data Analysis Software.

 

The CLARIOstarPlus has an internal barcode readers and is extremely fast and sensitive andis used primarily for high-throughput screens and platform specific applications (FP, TRF, HTRF), but can also be used for simple plate based absorbance reads and luminescence reads. The CLARIOstarPlus is integrated into the HTSKC main robotic HTS system and therefore it's stand-alone usage will require pre-approval so as not to interfere with screening operations. For independent use, use the instrument in CCSR-0133.

HTSKC BMG CLARIOstar Plus Technical Information (AlphaScreen/LISA only in CCSR)

Filter Description Wavelength-range Dichrioc Description Wavelength
UV-TR-Ex 295-10 UV LP 309
TR/Eu-HTRF-Ex 330-75 Alexa350 LP 410
UV-TR-Em 340-10 HTRF Eu LP TR (LP 426)
Coumarin-FP-Ex 360-20 FITC LP 504
Alexa350-FP-Em 450-10 Cy3 LP 566
NanoBRET-Em 450-80 TAMRA LP 601
FITC-FP-Ex 482-16 Cy5 LP 639
Terbium-HTRF-Em 490-10 Alpha* SP AS
FITC-FP-Em 520-10    
TAMRA-FP-Ex 530-40    
Texas Red-FP-Ex 540-20    
Alexa588-FP-Ex 580-10    
Cy3-FP-Em 590-20    
Texas Red-FP-Em 590-50    
TAMRA-FP-Em LP 610    
Alpha-EM* 615-18    
Eu-Em 620-10    
Chelate-Em 620-10    
Cy5-HTRF-Em 665-10    
Cy5-FP-Em 675-50    
Alpha-Ex* 680-40 x x

Infinite M1000 and Infinite M1000 PRO (Tecan) CCSR-0133

Tecan Infinite M1000

The Tecan Infinite M1000 and M1000 PRO are very flexible Quad4 monochromator-based 6 to 1536-well microplate reader and can measure;

  • Absorbance (230 nm-1,000 nm)
  • Fluorescence Intensity (Ex 230nm-850 nm, Em 280 nm-850 nm)
  • Fluorescence Polarization (FP, Ex (LED): 470, 530, 590, 635 nm)
  • Time-resolved fluorescence (TRF)
  • Fluorescence resonance energy transfer (FRET)
  • Homogeneous Time-Resolved Fluorescence (HTRF)
  • Glow/Flash Chemiluminescence with Dual Injectors for both 96 and 384-well plates.
  • Software data acquisition and analysis with Tecan Magellan or iControl.

 

Both Tecans have Barcode Readers and are extremely fast and sensitive and can be used primarily for high-throughput screens and platform specific applications (FP, TRF, HTRF), but can also be used for simple plate based absorbance reads and injector based luminescence reads.

The HTSKC Tecan Infinite M1000 PRO also has 50-microplate stackers and the AlphaScreen module .

Flexstation II 384 (Molecular Devices) CCSR-0133

Flexstation II 384

The FLEXstation II 384 (Molecular Devices, Inc.-Flex3 link -we don't have this) complements the Tecans, in that it has an internal 8- or 16-tip pipetting head and performs multi-label measurements [fluorescence, fluorescence resonance energy transfer (FRET) and chemiluminescence] in real-time in 96 or 384 well format. This enhanced capability allows for kinetic studies of biological events, such as G protein-coupled receptor activation (via Calcium mobilization, ie FLIPR screens) and ion channel gating in live cells. The FLEXstation also can read 6-, and 24-well plates for applications such as measuring cell migration. The FLEXstation is used primarily for platform specific applications, assay development and medium throughput screens. Data acquisition and analysis is performed with Molecular Devices Softmax Pro.

AnalystGT (Molecular Devices) CCSR-0128--Inactive

AnalystGT

The AnalystGT is very flexible filter-based 96-, 384-, and 1536-well microplate reader and can measure (see below for list of filters);

  • Absorbance
  • Fluorescence Intensity
  • Fluorescence Polarization (FP)
  • Time-resolved fluorescence (TRF)
  • Fluorescence resonance energy transfer (FRET)
  • Homogeneous Time-Resolved Fluorescence (HTRF)
  • Chemiluminescence

 

The HTSKC AnalystGT has 40-microplate stackers to allow its independent use. It is extremely fast and sensitive and was used primarily for high-throughput screens in conjection with the Caliper LS Staccator system and platform specific applications (FP, TRF, HTRF)

HTSKC AnalystGT Technical Information

Filter Description Wavelength-range Dichrioc Description Wavelength
UV DNA 260-Ex (2) 260-10 UV Dichrioc 380
UV Protein 280-Ex 280-10 UV Dichrioc 380
Cryptate-Ex (4) 330-70 UV Dichrioc 380
Chelate-Ex 330-80   400
Europium/Coumarin-Ex (6) 360-35   425
CCF2 Excitation (7) 405-20   425
Coumarin-Em (4) 425-35   490/550
CCF2 Donor Emission (3) 460-40 Fluorescein 505
Fluorescein-Ex (1) 485-20   505/600
CCF2 Acc Em, CellTiterBlueEx (5) 530-10   561
Fluorescein-Em (1) 530-25   650
Rhodamine-Em 580-10 50/50 Beamsplitter  
Texas Red-Ex (8) 580-10    
Cell-Titer Blue-Em (5) 590-20    
Cryptate Donor Emission (6) 615    
Chelate Donor Emission 620    
Texas Red-Em (7) 630-35    
Cy5/Dylight Excitation (3) 630-35    
Cryp/Chelate Acceptor Emission (8) 665-7.5    
Cy5/Dylight Emission (2) 682-22    
EPI-Absorbance Attenuator ND3    
Note that all filters can be used for either emission or excitation with any dichrioc. Parentheses indicate current position in Emission or Excitation wheels.