Clinical quality of breath-held T1-weighted breast MRI in the supine position

To enable supine breast MRI as an alternative to prone positioning, we need to revisit breast-specific image acquisition approaches to ensure we can provide the radiologist with images that are interpretable and of sufficient diagnostic quality for the specific indication. A T1-weighted imaging sequence that is run post contrast injection is the key component of every breast MRI protocol. 

This study assessed the clinical quality of T1-weighted 3D LAVA Flex for supine breast MRI. The sequence is deployed during a 19-second breath-hold to minimize motion artifacts. Dixon fat-water separation is used for robust suppression of fat signal. We employ a 60-channel body-conforming RF coil that was specifically designed for breast imaging in the supine position. We obtained feedback on image quality via a reader study involving three breast radiologists, particularly concerning breast coverage, sharpness, signal-to-noise-ratio, and the existance of residual artifacts.

Results of the study suggest that breath-held T1-weighted 3D LAVA Flex permits full bilateral supine breast MRI, that the body-conforming high-density coil allows for fast, highly-accelerated imaging, and overall underlines the potential of supine breast MRI as an alternative to the standard prone positioning.

Zimmermann J, Vincent J, Poplack SP, Pittman SM, Dashevsky BZ, Zhou X, Robb F, Daniel BL, Hargreaves BA, Clinical quality of breath-held T1-weighted breast MRI in the supine position, European Journal of Radiology 202 (2026) 112938, doi: https://doi.org/10.1016/j.ejrad.2026.112938

Online Journal Article

Image examples of (a–d) supine LAVA with R = 4 × 2 and (e–f) prone DISCO. LAVA images were acquired at ≥21 min post-contrast. Prone images show the final delayed temporal phase (≥7 min post contrast). Supine images (a, b, c) were rated ‘‘of diagnostic quality’’ by all readers. Image (a) was rated ‘equal SoC’ or ‘exceeding SoC’ (all categories), while image (b, c) show loss in sharpness and/or presence of artifacts (green arrows), leading to ‘below SoC’ (yet still diagnostic) quality. Image (d), patient with large body habitus, was unanimously rated ‘‘insufficient quality for diagnostic usage’’ owing to right breast SNR loss due to missing coil coverage, and not enough right-left field of view (45 cm, orange arrow). SoC: standard-of-care.