  • Efficient Bloch-Siegert B(1) (+) mapping using spiral and echo-planar readouts.

    Saranathan M, Khalighi MM, Glover GH, Pandit P, Rutt BK. Magn Reson Med. 2013

    The Bloch-Siegert (B-S) B(1) (+) mapping technique is a fast, phase-based method that is highly SAR limited especially at 7T, necessitating the use of long repetition times. Spiral and echo-planar readouts were incorporated in a gradient-echo based B-S sequence to reduce specific absoprtion rate (SAR) and improve its scan efficiency. A novel, numerically optimized 4 ms B-S off-resonant pulse at + 1960 Hz was used to increase sensitivity and further reduce SAR compared with the conventional 6 ms Fermi B-S pulse. Using echo-planar and spiral readouts, scan time reductions of 8-16 were achieved. By reducing the B-S pulse width by a factor of 1.5, SAR was reduced by a factor of 1.5 and overall sensitivity was increased by a factor of 1.33 due to the nearly halved resonance offset of the new B-S pulse. This was validated on phantoms and volunteers at 7 T. Magn Reson Med, 2013. © 2013 Wiley Periodicals, Inc.