Radiology

Huy M. Do

Publication Details

  • Delayed Retraction of the Pipeline Embolization Device and Corking Failure: Pitfalls of PED Placement in the Setting of a Ruptured Aneurysm.

    McTaggart RA, Santarelli JG, Marcellus ML, Steinberg GK, Dodd RL, Do HM, Marks MP. Neurosurgery. 2012

    BACKGROUND AND IMPORTANCE:: The safety of flow-diverting stents for the treatment of ruptured intracranial aneurysms is unknown. CLINICAL PRESENTATION:: A 35-year-old woman with a ruptured, dissecting aneurysm of the intradural right vertebral artery and incorporating the right posterior inferior cerebellar artery was treated with a Pipeline Embolization Device (PED). Five days after reconstruction of the diseased right vertebral segment, she was treated for vasospasm, and retraction of the PED was observed, leaving her dissecting aneurysm unprotected. A second PED was placed with coverage of the aneurysm, but vasospasm complicated optimal positioning of the device. CONCLUSION:: In addition to the potential risks of dual-antiplatelet therapy in these patients, this case illustrates two pitfalls of flow-diverting devices in vessels in vasospasm: delayed retraction of the device and difficulty positioning the device for deployment in the setting of vasospasm.

    PubMedID: 23190640