  • Dual-axis confocal microscope for high-resolution in vivo imaging.

    Wang TD, Mandella MJ, Contag CH, Kino GS. Opt Lett. 2003; 28 (6): 414-6

    We describe a novel confocal microscope that uses separate low-numerical-aperture objectives with the illumination and collection axes crossed at angle theta from the midline. This architecture collects images in scattering media with high transverse and axial resolution, long working distance, large field of view, and reduced noise from scattered light. We measured transverse and axial (FWHM) resolution of 1.3 and 2.1 microm, respectively, in free space, and confirm subcellular resolution in excised esophageal mucosa. The optics may be scaled to millimeter dimensions and fiber coupled for collection of high-resolution images in vivo.

    PubMedID: 12659264