Community Academic Profiles

Michael V. McConnell, MD, MSEE

Publication Details

  • Hybrid referenceless and multibaseline subtraction MR thermometry for monitoring thermal therapies in moving organs.

    Grissom WA, Rieke V, Holbrook AB, Medan Y, Lustig M, Santos J, McConnell MV, Pauly KB. Med Phys. 2010; 37 (9): 5014-26

    Magnetic resonance thermometry using the proton resonance frequency (PRF) shift is a promising technique for guiding thermal ablation. For temperature monitoring in moving organs, such as the liver and the heart, problems with motion must be addressed. Multi-baseline subtraction techniques have been proposed, which use a library of baseline images covering the respiratory and cardiac cycle. However, main field shifts due to lung and diaphragm motion can cause large inaccuracies in multi-baseline subtraction. Referenceless thermometry methods based on polynomial phase regression are immune to motion and susceptibility shifts. While referenceless methods can accurately estimate temperature within the organ, in general, the background phase at organ/tissue interfaces requires large polynomial orders to fit, leading to increased danger that the heated region itself will be fitted by the polynomial and thermal dose will be underestimated. In this paper, a hybrid method for PRF thermometry in moving organs is presented that combines the strengths of referenceless and multi-baseline thermometry.

    PubMedID: 20964221

Stanford Medicine Resources:

Footer Links: