Cancer Institute A national cancer institute
designated cancer center

Robert Herfkens

Publication Details

  • Steady-state free precession MR imaging: improved myocardial tag persistence and signal-to-noise ratio for analysis of myocardial motion.

    Markl M, Reeder SB, Chan FP, Alley MT, Herfkens RJ, Pelc NJ. Radiology. 2004; 230 (3): 852-61

    Tagging with balanced steady-state free-precession (SSFP) magnetic resonance (MR) imaging by using a steady-state storage scheme for myocardial motion analysis was evaluated. Signal-to-noise ratio (SNR), blood-tissue contrast, and tag persistence in volunteers and phantoms showed improved performance of SSFP imaging with tagging compared with that of radiofrequency spoiled gradient-echo (SPGR) MR imaging with tagging. Choice of flip angle with SSFP imaging involved a trade-off among SNR, blood-tissue contrast, and tag persistence. Increased SNR and tag persistence can be achieved simultaneously with SSFP imaging compared with SPGR tagging methods. As a result, the proposed technique may be useful for analysis of diastolic ventricular function.

    PubMedID: 14990847

Stanford Medicine Resources:

Footer Links: